EasyManua.ls Logo

GE Vivid 4 BT03 RFI - VIVID 4 17 Monitor Replacement Procedure; Preparation; Tools; VIVID 4 17 Monitor Removal

GE Vivid 4 BT03 RFI
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-24 Section 8-3 - Control Console Components Replacement
8-3-2 Vivid™ 4 17" Monitor Replacement Procedure
CAUTION: The Vivid™ 4 17" monitor is heavy. At least two people must be available to lift the
monitor. Attempts to move this unit by one person alone, could result in personal injury, and/or
damage to the system.
8-3-2-1 Tools
Use the appropriate Phillips-type screw drivers as indicated in the Vivid™ 4 monitor replacement
procedures.
8-3-2-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-3-2-3 Vivid™ 4 17" Monitor Removal
1.) Turn OFF the Vivid™ 4 main circuit breaker, located at the rear of the machine (lower end).
2.) Disconnect the main AC power cable from the wall outlet.
3.) Remove any Color printer (if present) from the side compartment.
4.) Remove the six screws from the metal cover on the base assembly and remove the cover, as
shown in Figure 8-22 below.
5.) Unplug the VGA cable and AC power cable from the receptacle cable inside the service
compartment (the metal cover was removed in step 4 to provide easy access).
Figure 8-22 Metal Cover Removed
Note: One screw hole not visible in picture

Table of Contents

Related product manuals