GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-102 Section 8-6 - Lower Section Components Replacement
8-6-7-4 Up/Down/Swivel Handle Installation Procedure
NOTE: This describes the procedure for installing the Up/Down/Swivel Handle (Part No. 237002).
1) Place the new Up/Down/Swivel handle into the correct position on the handle mounting brackets 
(fastened to the machine chassis) - refer to Figure 8-89 on page 8-101.
2) Return the M6 Axis screw (previously removed) passing the end through the screw hole in the first 
handle mounting bracket, through the two holes in the handle, then through the screw hole in the 
second bracket - refer to Figure 8-89 on page 8-101.
3.) Return the M6 nut (previously removed) to the handle assembly axis screw. Tighten the nut fully to 
secure the handle assembly to the handle mounting brackets - refer to Figure 8-89 on page 8-101.
4.) Insert the gas spring cable end stopper through the grooved slot on the handle assembly carefully 
aligning the groove on the cable with that of the groove in the slot (refer to the detail shown in Figure 
8-88 on page 8-101).
5.) Feed the gas spring cable down into the slot in the cable retaining bracket (refer to Figure 8-90 
below) and tighten the securing nut against the bracket to hold the cable firmly in position - refer 
also to Figure 8-87 on page 8-100.
6.) Repeat step 4 to insert the swivel release cable end stopper into the grooved slot on the handle 
assembly.
7.) Repeat step 5 to secure the swivel release cable to the cable retaining bracket (Figure 8-90 below).
8.) Slide the cable retaining bracket backwards (previously moved forwards) along the chassis beam 
sufficiently to regain tension on the cables. Using an M4 Allen key, tighten the two cap screws on 
the bracket to hold it firmly in position - refer to Figure 8-85 on page 8-99.
9) Test the Up/Down/Swivel handle to make sure the mechanism is operating properly and that the 
handle is fastened securely on the chassis.
10.)Return the keyboard (previously removed) as described in the Keyboard Installation Procedure on 
page 8 - 27.
11.)Reconnect the four keyboard matrix cables - refer to Figure 8-83 on page 8-98.
12.)Return the control console upper cover front section, as described in the Control Console Upper 
Cover (Front) Installation Procedure on page 8 - 15.
13.)Return the control console lower cover, as described in the Control Console Lower Cover 
Installation Procedure on page 8 - 18.
14.)Return the control console lower cover rear, as described in the Control Console Lower Rear Cover 
Installation Procedure on page 8 - 19. 
Figure 8-90 Swivel Release Cable Secured in Retaining Bracket
Cable securing nut
Cable retaining bracket