GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-100 Section 8-6 - Lower Section Components Replacement
Note: For clarification, Figure 8-86 below shows a close-up view of the up/down/swivel handle, 
gas spring cable, and swivel release cable fitted in position, as seen from underneath .
7.) Using a combination wrench, release the swivel release cable from the cable retaining bracket by 
loosening the securing nut, as shown in Figure 8-87 below. 
8.) Release the swivel release cable from the up/down/swivel handle by carefully sliding the cable end 
stopper out through the grooved slot on the handle assembly (refer to the detail shown in 
Figure 8-88 on page 8-101).
9.) Repeat step 7 to release the gas spring cable from the cable retaining bracket.
10.)Repeat step 8 to release the gas spring cable from the up/down/swivel handle (refer to the detail 
shown in Figure 8-88 on page 8-101).
Figure 8-86 Handle and Gas Spring Cables - Close-up View
Figure 8-87 Releasing the Gas Spring Cable from the Cable Retaining Bracket
Cable retaining bracket
Handle
Swivel release cable
Cable securing nut
Gas spring cable
Swivel release cable