7-48 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9 MUX Diagnostic Tests
NOTE: The Select Socket and Select Supply options are not relevant for this test.
7-3-9-1 Accessing the MUX Diagnostic Test Options
1) Access the FE Diagnostics dialog box, as described in the Accessing the Front End Diagnostic
Options section, on page 7-6.
2) In the FE Diagnostics dialog box, trackball to the MUX tab and press Select,
OR
Trackball to the MUX button in the Data Flow Maps and press Select. The MUX tab is displayed,
as shown in Figure 7-26 (or Figure 7-27) below:
Figure 7-26 MUX Tab - FE Diagnostics Dialog Box (RFI Configuration)
Data Flow Map
Test Description Area
Test Area
Comment Area
Tabs