GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-114 Section 8-8 - Peripherals
Section 8-8
Peripherals
8-8-1 B/W Video Printer Replacement Procedure
NOTE: This section describes the installation and removal procedure for the Sony UP-895MDW B/W Video
Printer.
Figure 8-100 SONY B/W Video Printer
8-8-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers.
8-8-1-2 Manpower
One person - 30 mins.
8-8-1-3 Preparations
1) Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panel cover as shown in the Connector Panels Cover Removal Procedure
on page 8 - 8.
3.) Unpack the B/W Video Printer.
4.) Check that the B/W Video Printer AC voltage corresponds to that of your local voltage.
5.) Verify that all the DIP Switches on the rear panel of the B/W Video Printer are in the Down position
as shown in Figure 8-101 on page 8-115.