GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-118 Section 8-8 - Peripherals
8-8-2 Mitsubishi VCR Replacement Procedure
NOTE: This section describes the installation and removal procedure of the following MITSUBISHI VCR models:
• HS-MD3000E (PAL) - P/N: 066E9513
• HS-MD3000U (NTSC) - P/N: 066E9514
The VCR will not operate without initial setup and adjustments, as described below.
Figure 8-105 Mitsubishi Video Cassette Recorder (VCR)
8-8-2-1 Tools
No special tools needed.
8-8-2-2 Manpower
One person - 30 mins.
8-8-2-3 Preparations
1) Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal 
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.