8-2-11 Control Console Lower Rear Cover Replacement Procedure
8-2-11-1 Tools
Use the appropriate Phillips-type screw drivers, as indicated in the replacement procedures of the upper
section of the control console’s bottom cover.
8-2-11-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-2-11-3 Control Console Lower Rear Cover Removal Procedure
1) Raise the control console to its maximum height.
2) Remove the 4 screws that fasten the rear cover in position (two at the rear, and two below) - see
Figure 8-18 below:
Figure 8-18 Location of Screws on Lower Rear Cover
3) Remove the bottom rear cover.
8-2-11-4 Control Console Lower Rear Cover Installation Procedure
1) Raise the control console to its maximum height.
2) Return the lower rear cover to its original position at the rear of the machine.
3) Fasten the lower cover with its four screws.