GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
2-6 Section 2-2 - Console Requirements
The following table lists recommendations for preventing EMI:
2-2-7 Probe Environmental Requirements
Note: System and electronic probes are designed for storage temperatures of -20
o
 to +50
o
 C. When 
exposed to large temperature variations, the probes should be kept at room temperature for 
a minimum of 10 hours before use. 
Table 2-4    EMI Prevention/ Abatement
EMI Rule Details
Ground the unit.
Poor grounding is the most likely reason an ultrasound unit will have noisy images. 
Check the grounding of the power cord and power outlet.
Be aware of RF sources.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may 
be required to eliminate interference problems caused by high frequency, high 
powered radio or video broadcast signals.
Replace and/or reassemble all 
screws, RF gaskets, covers and 
cores.
After you finish repairing or updating the system, replace all covers and tighten all 
screws. Any cable with an external connection requires a magnet wrap at each end. 
Install the shield over the front of the card cage. Loose or missing covers or RF gaskets 
allow radio frequencies to interfere with the ultrasound signals.
Replace broken RF gaskets.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the 
gasket. Do not turn ON the unit until any loose metallic part is removed and replaced, 
if required.
Do not place labels where 
RF gaskets touch metal.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will 
permit RF leakage. In case a label has been found in such a location, move the label 
to a different, appropriate location.
Use GE specified harnesses and 
peripherals.
The interconnect cables are grounded and require ferrite beads and other shielding. 
Cable length, material, and routing are all important; do not make any changes that do 
not meet all specifications.
Take care with cellular phones. Cellular phones may transmit a 5 V/m signal that causes image artifacts. 
Properly address peripheral cables.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral 
bays. Loop any peripheral cable excess length inside the peripheral bays or hang on 
the hooks provided below the console. Attach the monitor cables to the frame.
Table 2-5    Probe Operation and Storage Temperatures
Electronics PAMPTE
Operation
10 — 40
o
C (50 — 104
o
F) 5 — 42.7
o
C (41 — 109
o
F)
Storage
-20 — 50
o
C (-4 — 122
o
F) -20 — 50
o
C (-4 — 122
o
F)