GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-126 Section 8-8 - Peripherals
8-8-3 Sony VCR Replacement Procedure
NOTE: This section describes the installation and removal procedure of the SONY SVO-9500MDP2 and
SVO-9500MD2.
The SVO-9500MDP2 (PAL) and SVO-9500MD2 (NTSC) are alike, except for the Color Signal system.
The installation procedure is the same for both models.
Figure 8-118 SONY Video Cassette Recorder (VCR)
8-8-3-1 Tools
No special tools needed.
8-8-3-2 Manpower
One person - 30 mins.
8-8-3-3 Preparations
1) Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR front panel are as shown in Figure 8-119 below.
If necessary, adjust accordingly.
Figure 8-119 Sony VCR - Switches Setting (Front Panel)
Monitor Mix
Not important.
(Determines what is sent to MONITOR OUT jack.
Has no effect on AUDIO OUT).
Audio Hi-Fi
Output High Fidelity audio to AUDIO OUT
CH-1/L and CH-2/R
Hi-Fi ON Record High Fidelity audio, if present
Ch1
Mix
Ch2
Monitor Audio
Hi-Fi
Hi-Fi
Mix
Norm
On
Off