8-88 Section 8-6 - Lower Section Components Replacement
8-6-4 AC, BEP or FE Cable Replacement Procedure
Use the appropriate flat and Phillips-type screw drivers, a wire cutter, the appropriate Hex wrenches
and the appropriate Allen keys as indicated in the cable replacement procedure.
8-6-4-1 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-6-4-2 AC, BEP or FE Cable Removal Procedure
1) Disconnect both ends of the required cable.
2) Open the plastic ties that hold the cables together.
3) Remove the required cable.
8-6-4-3 AC, BEP or FE Cable Installation Procedure
1) Route the required cable.
2) Connect both ends of the new cable.
3) Replace the ties and secure the cable in position.