4-6 Section 4-3 - Functional Check
4-3-3 Mechanical Functions
4-3-3-1 Console Movement Test
4-3-3-2 Brakes Function Test
Table 4-11 Console Movement Test
Step Task Expected Result(s)
1
Pull the front handle (located under the control console) to
the half-pressed position and swivel the console to the left.
Ensure that you do not apply too much force to move
the console and that it moves smoothly to the left.
2
With the front handle still in the half-pressed position,
swivel the console to the right.
Ensure that you do not apply too much force to move
the console and that it moves smoothly to the right.
3
Pull the front handle to the fully-pressed position and lift
the console to its maximum height.
Ensure that you do not apply too much force to move
the console and that the console movement is
smooth.
4
With the front handle still in the fully-pressed position, push
the console to its minimum height.
Ensure that you do not apply too much force to move
the console and that the console movement is
smooth.
Table 4-12 Brakes Function Test
Step Task Expected Result(s)
1
Push the brake pedal to its center position. Push and pull
the unit right, left, backwards and forwards.
Ensure that the wheels move freely in all directions.
Check the wheels for wear and tear, and replace if
necessary.
2
Push the brake pedal to its left position (parking). Push
and pull the unit right, left, backwards and forwards.
Ensure that the wheels are locked and there is no
movement in any direction.
3
Push the brake pedal to its right position (arrows). Push
and pull the unit right, left, backwards and forwards.
Ensure that the front wheels are locked for right and left
movement only, but can move freely backwards and
forwards.