IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 4 - Functional Checks 4-5
4-3-2-2 Video OUT Test
4-3-2-3 B/W Printer Tests
4-3-2-4 Color Printer Tests
Table 4-8 Video OUT Test
Step Task Expected Result(s)
1
Activate a CineLoop or Freeze a scanned image.
2
Connect the monitor to the composite VIDEO OUT 1
connector and check the image.
The image is displayed correctly.
3
Connect the monitor to the composite VIDEO OUT 2
connector and check the image.
The image is displayed correctly.
4
Connect the monitor to the composite VIDEO B/W
connector and check the image.
See B/W Printer Tests, below.
The image is displayed correctly.
5
Connect the monitor to the Color Printer RGB
connectors and check the image.
See Color Printer Tests, below.
The image is displayed correctly.
Table 4-9 B/W Printer Test
Step Task Expected Result(s)
1
Scan a phantom or suitable interface in 2D-Mode.
2
Press the Freeze button.
3
Press the Print B button.
Evaluate the print quality. Adjust the brightness and/or
contrast of the B/W printer, if required.
Table 4-10 Color Printer Test
Step Task Expected Result(s)
1
Scan a blood vessels in 2D-Mode or a phantom
simulating a similar CFM effect, then activate CFM to
see color.
2
Press the Freeze button.
3
Press the Print A button.
Evaluate the color print quality. Adjust the brightness
and/or contrast of the Color printer, if required.