GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-59
6) The battery is fastened with velcro strips to the inner wall of the BEP. Remove the battery by pulling
it away from the velcro.
Figure 8-49 Plug & Scan Battery Removed
8-5-5-4 Plug & Scan Battery Installation Procedure
Note: When replacing the battery, first pull off and remove the top layer from the two velcro
fastening strips on the rear of the battery, as shown below. These strips are not required
because two similar strips from the old battery are already found on the inner wall of the BEP.
Figure 8-50 Removing Plug & Scan Velcro Strips
1) Place the battery inside the battery bracket. Ensure that the cable is positioned at the lower end and
that the rear of the battery (with the two velcro strips) faces the inner wall of the BEP.
2) Mount the battery and bracket in its original location in the BEP, tightening the screws (two or three
as shown in Figure 8-48 on page 8-58) to secure it firmly in position.
3) Reconnect the battery cable to the Plug & Scan board.
4) Mount the BEP side cover as described in BEP2 - Side Cover Installation Procedure on page 8 - 54.
5) Reconnect the power cable to the rear of the BEP.
6) Return the left cover, as described in the Side Covers Installation Procedure on page 8 - 4.
Note: Perform a complete functional check of the Plug & Scan (see Chapter 4 - Functional
Checks).