GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
7-106 Section 7-5 - Common Service Interface
7-5-4-3-9 Audio Test
The audio test option is not enabled.
7-5-4-3-10 Monitor Test
When PC Diagnostics is selected in the left pane of the Diagnostics page, the following monitor
diagnostic tests can be performed, as shown below:
Figure 7-80 Monitor Test
Combination Test: Checks the monitor alignment, color depth and monitor resolution. Four
cross hairs are placed in each corner of the screen, as well as RGB spectrum arrays and
horizontal and vertical alignment bars.
Solid Color Test: Tests the panel display on laptops and LCD screens. This test display a solid
color on the screen to ensure that all the color pixels are operating correctly.
VESA (Video Electronics Standards Association) Test Patterns: Checks luminance,
geometry and focus. These tests are designed for professional measurement use.

Table of Contents