EasyManua.ls Logo

GE Vivid 4 BT03 RFI - Section 4-4 Image Testing: 2 D;M;CFM;Doppler

GE Vivid 4 BT03 RFI
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 4 - Functional Checks 4-11
Section 4-4
Image Testing: 2D/M/CFM/Doppler
NOTE: To perform these tests thoroughly, it is recommended to use three types of soft tissue phantoms: RMI
403GS, RMI 404G and RMI 1425A, or an equivalent surface that can simulate the same soft tissue
effects.
4-4-1 3S Probe Image Quality Tests
4-4-1-1 2D Noise Uniformity Test
4-4-1-2 2D Performance Test
Table 4-18 2D Noise Uniformity Test (3S)
Step Task Expected Result(s)
1
With a 3S probe in the air, select a
Cardiac preset.
2
Set the following parameters:
•2D Gain = 50
Depth = 7cm
Dynamic Range = 80dB
Ensure that there are no white circular lines or other patterns in
the noise.
Table 4-19 2D Performance Test (3S)
Step Task Expected Result(s)
1
Use the Standard Imaging Phantom
RMI403GS.
2
Select a Cardiac preset.
3
Set the TGC so that the image is uniform.

Table of Contents

Related product manuals