Chapter 7 - Diagnostics/Troubleshooting  7-59
7-3-9-6 Advanced Board Tests
The advanced board tests check two additional parts of the MUX Board: TEE probe control and 
Monitoring, as follows:
- TEE probe control enables you to check different positions of the probe; by changing the 
probe angle, you are able to verify its normal operation. 
- Monitoring enabled you to check the HVPS and LVPS voltages, and the AC Power Input, as 
well as temperatures.
7-3-9-6-1 TEE Probe Control
The TEE probe control test checks the position setting of the probe, as well as its normal operation.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the Accessing the MUX 
Diagnostic Test Options section, on page 7-48.
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests 
options are displayed, as shown in Figure 7-34 on page 7-59.
Figure 7-34 MUX Advanced System Tests Options
3) Select the radio button next to the appropriate socket.
4) Trackball to the TEE Probe Control button and press Select. The TEE Probe Test dialog box is 
displayed, as shown in Figure 7-35, with all the probe settings.