GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures  8-129
6.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation 
Procedure on page 8 - 8.
7) Reconnect the mains AC power cable from the wall outlet.
8) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
9) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
10.)Under System Configuration, VCR/ECG Tab, select the appropriate VCR mode options. For more 
information, see VCR/ECG Tab on page 3 - 45. 
Note: Sony VCR - Reset Counter:
If the Reset Counter button is activated while the VCR is in the Stop, Pause, or Record 
state, the counter display is reset to zero (0:00:00).
8-8-3-5 Sony VCR Removal Procedure
1.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal 
Procedure on page 8 - 8.
2.) Release the B/W printer by undoing the fastening belt around it (together with the VCR, if present).
3.) Disconnect the RS-232C cable, the Audio and S-Video IN and OUT cables, and Power cord from 
VCR’s rear as shown in Figure 8-122 on page 8-128. 
4.) Remove the VCR from the left side of the Vivid™ 4 ultrasound unit (taking it out from below the 
control console and from beside the B/W printer, if present) - refer to Figure 8-121 on page 8-127.