IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-119
8-8-2-4 Mitsubishi VCR Installation Procedure
1.) Place the VCR on the left side of the Vivid™ 4 ultrasound unit’s top surface (below the control
console - as shown in Figure 8-106 below) and next to the B/W Printer, if present, allowing sufficient
room to access the cables at the rear of the VCR.
2.) Connect one end of the VCR power cable (P/N: 2269430) to the power outlet on the Vivid™ 4 rear
panel and the other end to the AC power socket on the VCR rear panel - see Figure 8-107 on
page 8-119.
3.) Using the S-video cable (P/N: 2266743), connect one end to the Video OUT socket on the VCR
rear panel (refer to Figure 8-107) and the other to the Video IN socket on the Vivid 3 rear panel.
4.) Boot up the Vivid™ 4 ultrasound unit by pressing the ON/OFF button on the control console.
Figure 8-106 Mitsubishi VCR Mounted on Vivid™ 4 Scanner
Figure 8-107 Mitsubishi VCR Rear Panel
Video Cassette
Recorder (VCR)
AC Power Socket
Video OUT Socket
Audio OUT Socket
Audio IN Socket
RS-232C Socket