8-18 Section 8-2 - Cover Replacement Procedures
8-2-10 Control Console Lower Cover Replacement Procedure
8-2-10-1 Tools
Use the appropriate Phillips-type screw drivers, as indicated in the bottom cover replacement
procedures.
8-2-10-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-2-10-3 Control Console Lower Cover Removal Procedure
1) Raise the control console monitor to its maximum height.
2) Remove the probe cable holders (one from each side).
3.) Remove the control console lower rear cover, as described in the Control Console Lower Rear
Cover Removal Procedure on page 8 - 19.
4) Remove the 16 screws (7 from each side and 2 from the front) that fasten the lower cover in place,
as shown in Figure 8-17 below.
Figure 8-17 Location of Screws on the Control Console Lower Cover
5) Lower the control console lower cover downwards.
6.) Completely remove the cover from the machine.
8-2-10-4 Control Console Lower Cover Installation Procedure
1) Return the control console lower cover to the machine, placing it in the correct position.
2) Replace the 16 screws (7 on each side and 2 in the front [previously removed]) and fasten the cover
in place.
3.) Return the control console lower rear cover, as described in the Control Console Lower Rear Cover
Installation Procedure on page 8 - 19.
4) Replace the probe cable holders (one on each side [previously removed]).