IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-98 Section 8-6 - Lower Section Components Replacement
8-6-7 Up/Down/Swivel Handle Replacement Procedure
NOTE: This section describes the removal and installation procedures for the metal up/down/swivel handle
(Part No. 237002).
8-6-7-1 Tools
Use the appropriate the Phillips-type screw drivers, Allen key, and combination wrench, as indicated in
the up/down/swivel handle removal procedure.
8-6-7-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-6-7-3 Up/Down/Swivel Handle Removal Procedure
1) Remove the control console lower cover rear, as described in the Control Console Lower Rear
Cover Removal Procedure on page 8 - 19.
2) Remove the control console lower cover, as described in the Control Console Lower Cover
Removal Procedure on page 8 - 18.
3) Remove the control console upper cover front section, as described in the Control Console Upper
Cover (Front) Removal Procedure on page 8 - 14.
4) Disconnect the four keyboard matrix cables as shown in Figure 8-83 below.
Figure 8-82 Vivid™ 4 Up/Down/Swivel Handle in Position
Figure 8-83 Disconnecting the Keyboard Matrix Cables
Up/Down/Swivel Handle
(used for control console
up/down/swivel
movement)