7-78 Section 7-4 - Performing Back End Diagnostics on the System
7-4-6 Media Driver Diagnostic Test
The media driver diagnostic test enables you to check the functionality of the MOD, if installed, as well 
as the functionality of the CD Writer, if installed.
7-4-6-1 MOD Diagnostic Test
WARNING: Do not run an MOD Diagnostic Test on MO media that is being used by the customer 
as a back-up device, since this test may cause unrecoverable damage to any backed-up patient 
information on the MO media.
1) Insert a new MO Media Sony EDM-2300B (2.3G) into the MOD.
1) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options 
section, on page 7-69.
2) Trackball to the Media Driver button and press Select. The Util_MO_CDR_App dialog box is 
displayed, as shown below:
Figure 7-52 Util_MO_CDR_App Dialog Box
3) Trackball to the MO radio button and press Select.
4) Trackball to the IsMediaAccess button an press Select. The message Media in drive G is 
accessible is displayed.
5) Trackball to the Eject button and press Select. The message Media is outside is displayed. Verify 
that the media has actually been ejected.
6) Trackball to the Load button and press Select. The message Media is inside is displayed. Verify 
that the media is actually inside.
7) Trackball to the Format button and press Select. The message Media in drive G was formatted 
successfully is displayed.
8) Trackball to the GetVolInfo button and press Select. The message Media in drive G has 
following properties: Media Label = MY_LABEL, Media File-System Name = FAT is displayed.
9) Trackball to the GetBtPSector button and press Select. The message The Bytes Per Sector 
value of media in drive G is 512 is displayed.
10) Trackball to the SetVolLabel1 button and press Select.The message Media Label set to 
TEST_LABEL1 is displayed.