8-20 Section 8-2 - Cover Replacement Procedures
8-2-12 Front Handle Replacement Procedure
8-2-12-1 Tools
Use the appropriate Phillips-type screw drivers and an 8mm wrench, to replace the front handle.
8-2-12-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-2-12-3 Front Handle Removal Procedure
1) Remove the control console lower cover, as described in the Control Console Lower Cover
Removal Procedure on page 8 - 18.
2) Remove the control console upper cover front, as described in the Control Console Upper Cover
(Front) Removal Procedure on page 8 - 14.
3.) Remove the two 8mm securing nuts (two on each side) and one nut located below the keyboard,
parallel to the hand rest location.
4) Remove the Front Handle.
8-2-12-4 Front Handle Installation Procedure
1) Return the Front Handle to its original position on the front of the machine.
2) Fasten each side of the Front Handle beneath the keyboard with its 8 mm securing nuts (five in
total).
3) Return the control console upper cover front to its original position, as described in the Control
Console Upper Cover (Front) Removal Procedure on page 8 - 14.
4) Return the control console lower cover to its original position, as described in the Control Console
Lower Cover Removal Procedure on page 8 - 18.
Figure 8-19 Front Handle