IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-47
Figure 7-25 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFT Configuration)
3) Check the test results and proceed as follows:
• If no errors were identified, proceed with the next test.
• If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
• If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-11, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in Chapter 5 - Components
and Function (Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
4) Reboot the system.