Chapter 8 - Replacement Procedures  8-133
5.) Place the VCR in its correct final position.
6.) Secure the VCR by fastening the belt around it, or together with the B/W Printer (if present).
7.) If more secure fastening is desired, install the optional shelf (P/N H45011CN), which can be ordered 
separately if the Color Printer option is not installed.
8.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation 
Procedure on page 8 - 8.
9) Reconnect the mains AC power cable from the wall outlet.
10) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
11) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
12.)Under System Configuration, VCR/ECG Tab, select the Panasonic VCR option, as shown in 
Figure 8-127 below. For more information, see . 
Note: Panasonic VCR - Reset Counter:
If the Reset Counter button is activated while the VCR is in the Stop state, the counter 
display is reset to zero (0:00:00). 
If activated in either the Pause or Record state, the display continues running (counter does 
not zero).
Figure 8-127 Panasonic VCR Option