GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
6-12 Section 6-7 - Calibration
2) Trackball to the Calibration button and press Select. The following calibration options are
displayed:
Figure 6-6 Calibration Options