Chapter 7 - Diagnostics/Troubleshooting 7-33
Figure 7-18 Report Dialog Box - FLA Jumper Test (Closed)
The Report dialog box contains the following information:
Test Result Message: Displays the test result.
FB Elements: Displays a graphic representation of the front board elements, as follows:
Each element is represented by a button (highlighted in green) that displays the element,
and the channel number that is associated with that element.
Each button is highlighted yellow when selected.
Each button is highlighted red if errors were detected for one or more elements.
Checkboxes: Enable you to manipulate bad channels:
Show Only Bad Channels: Select this checkbox to display only bad channels in the
Display area.
Mark bad channel: Select this checkbox to manually mark the currently selected channel
as bad.