GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
7-82 Section 7-4 - Performing Back End Diagnostics on the System
Figure 7-55 Disk/Bus Bench Window
Disk Bench: In the information displayed, make sure that the speed values displayed are greater than
23MB/sec. This test checks the speed of information transfer between the hard disk and IDE BUS.
Bus Bench: In the information displayed, make sure that the speed values displayed are greater than
86MB/sec. This test checks the speed of information transfer between the memory and the BUS.