GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
7-116 Section 7-5 - Common Service Interface
• Clean UserDefs - used to clean all user-defined settings and restore the default
settings
Figure 7-90 Clear User-defined Settings
CAUTION: THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a
complete backup of the current settings if they will be required in the future.
When selecting Clean Userdefs, the following warning message is displayed:
After making a back-up copy of the current settings, click OK to proceed.
NOTE: This function performs the same procedure as that of Restore Factory Presets (accessed under the
System Configuration window - System tab).