8-46 Section 8-5 - Back End Parts Replacement
NOTE: IMPORTANT - When replacing the BEP, the new Back End Processor will be supplied in a basic
configuration with only the following components ready installed:
• CD-R driver
• Floppy driver
•HD driver
• Cables and Brackets
• Memory 512 MB
Therefore, the FE will need to remove some components from the removed BEP and install them in the
new BEP (as part of the BEP Installation procedure - described below).
CAUTION: Failure to follow this exact instruction will cause irreversible damage!
Figure 8-39 Back End Component Locations - Systems with RFI Configuration