IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
10-4 Section 10-3 - Periodic Maintenance Schedule
Table 10-2 Periodic Maintenance Schedule
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •* * or before each use
Clean Probe Holders
Clean Air Filter
More frequently if necessary, depending on
the environment
Inspect AC Mains Cable
*• *Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Clean Monitor
Inspect Wheels, Casters, Brakes and
Swivel Locks
*•* Mobile Unit Check Daily
Check Control Console Movement
*•* Mobile Unit Check Daily
Console
Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Peripheral
Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Surface Probe
Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Endocavity Probe (vaginal or anal)
Current Leakage Checks
As prescribed in Probe Manual.
Transesphongeal (TEE) Probe
Current Leakage Checks
As prescribed in Probe Manual.
Surgical Probe
Current Leakage Checks
As prescribed in Probe Manual.
Measurement Accuracy Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Probe/Phantom Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Functional Checks
Also after corrective maintenance,
or as required by your facility’s QA program.