IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 10 - Periodic Maintenance 10-11
NOTE: For convenience (or if the air filter is excessively dirty), replacement filters are available. Refer
to Chapter 9 - Renewal Parts for the air filter replacement part number.
Table 10-9 Air Filter Cleaning
Step Item Description
1
Air Filter
Remove the air filter located at the front of the scanner, above the probe connector sockets.
Clean the air filter as follows: shake it in an area away from the system, wash with a mild
soapy solution, rinse, and air dry.
2
Air Filter
Install the clean air filter.
Refer to Chapter 8 - Replacement Procedures for air filter installation instructions.