10-26 Section 10-3 - Periodic Maintenance Schedule
10-7-8-1 Data Sheet for ECG Current Leakage
The test passes when all readings measure less than the values shown in the tables below. Record all 
data on the Vivid™ 4 Preventative Maintenance Inspection Certificate (see page 10-31).
Table 10-18 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
AC Power 
Source
Maximum
Allowance Limit
GROUND 
OPEN
GROUND 
CLOSED
Patient Lead-to-Ground Current Leakage Test
and
Patient Lead-to-Lead Current Leakage Test
115V 10uA 10uA
220/240V 500uA 10uA
Table 10-19   Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power 
Source
Maximum
Allowance Limit
Patient Lead Isolation Current Test
115V 20uA
220/240V 5mA
Table 10-20 Typical Data Sheet for ECG Lead-to-Lead Test
ECG 
Power
Tester 
Polarity 
Switch
Tester 
Ground 
Switch
Tester Lead Selector
RL
(N)
RA
(R)
LA
(L) LL C
ON NORM CLOSED
ON REVERSE CLOSED
ON NORM OPEN
ON REVERSE OPEN
OFF NORM CLOSED
OFF REVERSE CLOSED
OFF NORM OPEN
OFF REVERSE OPEN