GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-69
2.) In the displayed menus select [Media] -> [NewDisk] and then press ENTER.
NOTE: If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 13.0.0 or higher, a new software specific “Permanent key” is required.
3.) The following warning message will be displayed.
4.) ONLY if you are absolutely sure, select YES
and then confirm with ENTER.
5.) Creating a new disk will take some time......
Figure 5-48 Media -> NewDisk
Figure 5-49 WARNING! WARNING! WARNING!
NOTICE
If you select YES, all data on the entire hard disk will be lost and can not be recovered!
This includes all stored images and patient data!