_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:CONF:WITH:CUPP <data>
<data>
0.1 to 120 (NR1 or NR2)
Response
syntax
OK
Function
Sets the upper-limit value for withstand voltage tests, in the READY state.
Example
Sets the upper-limit value for withstand voltage tests.
Transmission :CONF:WITH:CUPP 5.0
The withstand voltage test upper-limit value is set to 5.0 mA.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Note that an execution error occurs when setting the test upper limit to a value less than the test
lower limit when the latter is on.
Syntax
:CONF:WITH:CUPP?
Response
syntax
<data>
<data>
0.1 to 120 (NR1 or NR2)
Function
Queries the upper-limit value for withstand voltage tests.
Example
Queries the upper-limit value for withstand voltage tests.
Transmission :CONF:WITH:CUPP?
Queries the withstand voltage test upper-limit value.
Response 5.0
The upper-limit value is 5.0 mA.
Syntax
:CONF:WITH:CLOW <data>
<data>
0.1 to 120 (NR1 or NR2)
Response
syntax
OK
Function
Sets the lower-limit value for withstand voltage tests, in the READY state.
Example
Sets the lower-limit value for withstand voltage tests.
Transmission :CONF:WITH:CLOW 0.1
The withstand voltage test lower-limit value is set to 0.1 mA.
Response OK
Settings completion
Error
The execution of this command in state other than the READY state causes an execution error.
Note that an execution error occurs when setting the test lower limit to a value more than the test
upper limit when the former is on.
:CONF:WITH:CUPP
Sets the upper-limit value for withstand voltage tests.
:CONF:WITH:CUPP?
Queries the upper-limit value for withstand voltage tests.
:CONF:WITH:CLOW
Sets the lower-limit value for withstand voltage tests.