9.4 Command
______________________________________________________________________________________________
Syntax
:CONF:WITH:CLOW?
Response
syntax
<data>
<data>
0.1 to 120 (NR1 or NR2)
Function
Queries the lower-limit value for withstand voltage tests.
Example
Queries the lower-limit value for withstand voltage tests.
Transmission :CONF:WITH:CLOW?
Queries the withstand voltage test lower-limit value.
Response 0.1
The lower-limit value is 0.1 mA.
Syntax
:CONF:WITH:TIM <data>
<data>
0.5 to 999 (NR1 or NR2)
Response
syntax
OK
Function
Sets the test time for withstand voltage tests, in the READY state.
Example
Sets the test time for withstand voltage tests.
Transmission :CONF:WITH:TIM 30.0
The withstand voltage test time is set to 30.0 s.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Syntax
:CONF:WITH:TIM?
Response
syntax
<data>
<data>
0.5 to 999 (NR1 or NR2)
Function
Queries the test time for withstand voltage tests.
Example
Queries the test time for withstand voltage tests.
Transmission :CONF:WITH:TIM?
Queries the withstand voltage test time.
Response 30.0
The test time is set to 30.0 s.
:CONF:WITH:CLOW?
Queries the lower-limit value for withstand voltage tests.
:CONF:WITH:TIM
Sets the test time for withstand voltage tests.
:CONF:WITH:TIM?
Queries the test time for withstand voltage tests.