9.4 Command
______________________________________________________________________________________________
Syntax
:CONF:INS:VOLT <data>
<data>
500,1000 (NR1)
Response
syntax
OK
Function
Sets the test voltage value for insulation resistance tests, in the READY state.
Example
Sets the test voltage value for insulation resistance tests.
Transmission :CONF:INS:VOLT 500
The insulation resistance test voltage value is set to 500 V.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Syntax
:CONF:INS:VOLT?
Response
syntax
<data>
<data>
500,1000 (NR1)
Function
Queries the test voltage value for insulation resistance tests.
Example
Queries the test voltage value for Insulation resistance tests.
Transmission :CONF:INS:VOLT?
Queries the insulation resistance test voltage value.
Response 500
The test voltage value is 500 V.
Syntax
:CONF:INS:RUPP <data>
<data>
0.20 to 2000 (NR1 or NR2)
Response
syntax
OK
Function
Sets the upper-limit value for insulation resistance tests, in the READY state.
Example
Sets the upper-limit value for insulation resistance tests.
Transmission :CONF:INS:RUPP 2000
The insulation resistance test upper-limit value is set to 2000 MΩ.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Note that an execution error occurs when setting the test upper limit to a value less than the test
lower limit when the former is on.
:CONF:INS:VOLT
Sets the test voltage value for insulation resistance tests.
:CONF:INS:VOLT?
Queries the test voltage value for insulation resistance tests.
:CONF:INS:RUPP
Sets the upper-limit value for insulation resistance tests.