140
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:CONF:INS:TIM <data>
<data>
0.5 to 999 (NR1 or NR2)
Response
syntax
OK
Function
Sets the test time for insulation resistance tests, in the READY state.
Example
Sets the test time for insulation resistance tests.
Transmission :CONF:INS:TIM 10.0
The insulation resistance test time is set to 10.0 s.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Syntax
:CONF:INS:TIM?
Response
syntax
<data>
<data>
0.5 to 999 (NR1 or NR2)
Function
Queries the test time for insulation resistance tests.
Example
Queries the test time for insulation resistance tests.
Transmission :CONF:INS:TIM?
Queries the insulation resistance test time.
Response 10.0
The test time is set to 10.0 s.
Syntax
:CONF:INS?
Response
syntax
<data>
<data>
Insulation resistance test voltage value: 500,1000 (NR1)
Upper-limit value : 0.2 to 2000 (NR1 or NR2)
Lower-limit value : 0.2 to 2000 (NR1 or NR2)
Test time value : 0.5 to 999 (NR1 or NR2)
Function
Queries the insulation resistance test settings.
Example
Queries the insulation resistance test settings.
Transmission :CONF:INS?
Queries the insulation resistance test settings.
Response 500,2000,10.0,10.0
Test voltage: 500 V; Upper-limit value: 2000 M;
Lower-limit value: 10.0 M; Test time value: 10.0 s
:CONF:INS:TIM
Sets the test time for insulation resistance tests.
:CONF:INS:TIM?
Queries the test time for insulation resistance tests.
:CONF:INS?
Queries the insulation resistance test settings.