141
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:STAR
Response
syntax
OK
Starts a test in the READY state.
Function
To start a test using this command, in the optional functions, set "START" for the RS command
to "1: Set."
Example
Starting a test
Transmission :STAR
Test Start
Response OK
Completion
Error
The execution of this command in a state other than the READY state causes an execution error.
In the optional functions, if the RS command START is set to "0: Not set," an execution error
will occur.
Note that an execution error occurs when double action is set to "1: ON" on the optional
function setting screen.
Syntax
:STOP
<data>
OK
Function
In the TEST state, forcibly ends a test.
Furthermore, releases the Hold function and returns to the READY state.
However, in the optional functions, when "FAIL Mode Function" is set to ON, the Hold function
cannot be disabled by this command.
Example
Forcibly ends a test.
Transmission :STOP
Forcibly ends a test.
Response OK
Completion
Error
In the optional functions, when "FAIL Mode Function" is set to ON, an execution error will
occur if the unit is in the FAIL Hold Mode.
:STAR
Starting a test.
:STOP
Forcibly ends a test and releases the hold state.