Performance Test Record (oscilloscope)
Test Settings Results
Voltage
Measurement
Accuracy
Channel 1
Channel 2
Zero Input
Zero Input
Limits
-13.7 V - -14.3 V
-3.43 V - -3.57 V
-1.37 V - -1.43 V
-137.0 mV - -143.0 mV
143.0 mV - 137.0 mV
1.43 V - 1.37 V
3.57 V - 3.43 V
14.3 V - 13.7 V
-13.7 V - -14.3 V
-3.43 V - -3.57 V
-1.37 V - -1.43 V
-137.0 mV - -143.0 mV
143.0 mV - 137.0 mV
1.43 V - 1.37 V
3.57 V - 3.43 V
14.3 V - 13.7 V
Measured
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
Offset Accuracy
Channel 1
Channel 2
Channel 1
Channel 2
Zero-input offset
4 V/Div
1 V/Div
100 mV/Div
10 mV/Div
4 V/Div
1 V/Div
100 mV/Div
10 mV/Div
DC-input offset
-35.00 V
-10.00 V
-2.00 V
+2.00 V
+10.00 V
+35.00 V
-35.00 V
-10.00 V
-2.00 V
+2.00 V
+10.00 V
+35.00 V
0.00 V± 320.0 mV
0.00 V± 80.0 mV
0.00 V± 8.0 mV
0.00 V± 800.0 µV
0.00 V± 320.0 mV
0.00 V± 80.0 mV
0.00 V± 8.0 mV
0.00 V± 800.0 µV
-35.4 V - -34.6 V
-10.1 V - -9.90 V
-2.02 V - -1.98 V
+1.98 V - +2.02 V
+9.90 V - +10.1 V
+34.6 V - +35.4 V
-35.4 V - -34.6 V
-10.1 V - -9.90 V
-2.02 V - -1.98 V
+1.98 V - +2.02 V
+9.90 V - +10.1 V
+34.6 V - +35.4 V
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________
________