Measurement and Display FunctionsMeasurement and Display Functions
Di splayed WaveformsDisplayed Waveforms 24 lines maximum, with scrolling across 96 waveforms.
Measurement FunctionsMeasurement Functions
Run/Stop FunctionsRun/Stop Functions Run starts acquisition of data in specified trace mode.
St opStop In single trace mode or the first run of a repetitive acquisition, Stop halts
acquisition and displays the current acquisition data. For subsequent runs in repetitive
mode, Stop halts acquisition of data and does not change the current display.
Trace ModeTrace Mode Single mode acquires data once per trace specification. Repetitive mode
repeats single mode acquisitions until Stop is pressed or until the time interval between
two specified patterns is less than or greater than a specified value, or within or not
within a specified range.
IndicatorsIndicators
Activity IndicatorsActivity Indicators Provided in the Configuration and Format menus for identifying
high, low, or changing states on the inputs.
MarkersMar k er s Two markers ( X and O) are shown as vert ical dashed lines on t he display.
Tr i ggerTrigger Displayed as a vertical dashed line in the Timing Waveform display and as line 0
in the State Listing display.
Data Entry/DisplayData Entry/Display
LabelsLabel s Channels may be grouped t ogether and given a 6-charact er name. Up to
126 labels in each analyzer may be assigned with up to 32 channels per label.
Display ModesDi spl ay M odes St at e list ing, St at e Waveforms, Chart , Compare List ing, Compare
Difference List ing, Timing Waveforms, and Timing List ings. St at e List ing and Timing
Waveforms can be time-correlated on the same displays.
Timing WaveformTiming Waveform Pattern readout of timing waveforms at X or O marker.
BasesBases Binary, Octal, Decimal, Hexadecimal, ASCII (display only), Twoās Complement,
and User-defined symbols.
SymbolsSymbols 1,000 maximum. Symbols can be downloaded over RS-232 or HP-IB.
General Information
Supplemental Characteristics (logic analyzer)
1ā7