HT2800T_en_c.doc Page 100 of 275
8. The autosampler will start “Plunger zero” to automatically complete syringe installation.
9. The autosampler will also automatically start the syringe conditioning. This procedure
should take about 90 minutes.
4.2.2.1 Syringe warmer assembly removal
To remove the syringe warmer assembly follow the instructions given in paragraph 4.2.2
Setup: Syringe warmer assembly installation/ replacement (Headspace mode)” excluding
points “d)“-“f)“. Instead of executing these points simply remove the syringe warmer assembly.
Warning
Plunger, syringe and needle are heated up to 150°C. We recommend taking care to
prevent injury.
Warning
The syringe conditioning procedure can be skipped during the autosampler
installation because the syringe has been already conditioned by the manufacturer.
This procedure can also be skipped in case of the same syringe re-installation. If
the syringe barrel or the syringe plunger or both the parts are new, the procedure
must be carried out.