HT2800T_en_c.doc Page 195 of 275
Warning
If syringeID system is used, be careful while handling the syringe holder of the SyringeID
system (1.7.2 “Options and 11.2 “Liquid mode). Do not touch the connector or the
electronic circuit placed on the back of the holder. The components may be damaged. If
not used, the SyringeID holder must be stored in the antistatic bag supplied with the
packaging.
position on the back of the autosampler (see paragraph 2. 1. 1”HT2800T/HT2850T
Overview “), then press “CONTINUE”.
Figure 169: Spacer removal
7) “Remove vial from oven” screen: remove the 20ml vial from C position of the oven.
8) If during the final Zero procedure the alignment on injector is not correct please modify
it (see paragraphs 4 . 2 . 1 . 2 ” Alignment: Front Injector” a n d
4 . 2 . 1 . 3 ” Alignment: Rear Injector“).
7.10.4 Change from Liquid to SPME mode