HT2800T_en_c.doc Page 146 of 275
5.2.3 Method tools
Access sequence: RUN>SETTINGS>METHODS>TOOLS
In the “METHOD TOOLS” screen the following options are available:
•
Restore Default: used to load the default settings for a method;
•
Copy Methods: used to copy the parameters from one method to another.
If you select “RESTORE DEFAULT”, you have to select the method to reset. A confirmation
screen appears “Loading default parameter for method X”. Tap “LOAD” to restore the default
parameters and “CANCEL” not to make the restore operation. If you select “COPY METHODS”,
you have to select the source method and the destination method. Tap “CONTINUE” to confirm
and “ABORT” not to make the method copy.
Select “Exit” to exit from this screen and return to the “RUN: METHODS” screen (see paragraph
5.2.2 “Method X: Setting method parameters”).
5.2.4 Example
Below are some method examples, use these to create your own method.
Please keep in mind example methods should be adapted depending on your analysis
technique, analyzer model, inlet type, etc. and therefore these may not be suitable for your
application.
5.2.4.1 Typical GC Method (Liquid mode)
General parameters
Syringe Volume
10 ul
Analysis Time
5
Fill Speed 1ul/sec
Viscosity Delay
5sec
Solvent wash
parameters
Pre Wash Cycle (s)
3
Pre Wash Volume
10ul
Pre Wash Mode
Inject
Post Wash Cycle (s)
3
Post Wash Volume
10ul
Post Wash Mode
Inject
Internal standard
Enable
No
Volume
/
Air Gap Mode
/
5
Set it according to the elution time of your analyte(s) in your GC condition