HT2800T_en_c.doc Page 215 of 275
Tap “EXIT” and/or:
•
edit the syringe volume set in the method;
•
replace the syringe with one that has the correct capacity;
•
change the syringe volume set during installation if this doesn't match the installed
syringe.
Restart the single injection.
8.9.2.2 Syringe mismatch
This error appears only in Liquid mode, if SyringeID system is enabled (see paragraphs 11
“Appendix A – Glossary“ and 4.2.6.8 “Setup configuration: SyringeID“).
Error message:
“Syringe mismatch! Execute Syringe Installation”
Causes:
At the beginning of a single injection, the syringeID system detects the presence of a syringe
different from the one detected during the syringe installation procedure (see paragraph
4.2.3Syringe installation/ replacement (liquid mode).
Solution:
Tap “ABORT” and perform the syringe installation procedure (see paragraph 4.2.3 “Syringe
installation/ replacement (liquid mode)”. Then restart the single injection.
8.9.3 Automatic run mismatch errors
If the method parameters in any part of an automatic run conflict with those defined in “setup”,
the unit will display an error message. The unit will show where the problem occurs. These
include the following:
8.9.3.1 Syringe Volume mismatch
Error message:
“Warning: Run Error Parameters”
“Step X: Syringe Volume mismatch in Method Y”
(Where “X” is the sequence step and “Y” is the method used)
Causes:
The syringe volume set during installation is different from the syringe volume set in the method
used.
Solution:
Tap “EXIT” and/or:
edit the syringe volume set in the method;
replace the syringe with one that has the correct capacity;