HT2800T_en_c.doc Page 246 of 275
4.2.2.1 “Syringe warmer assembly removal”. Remove any residual substances from the
instrument in accordance with the applicable safety regulations;
2. Remove sample vials and dispose of any residual substances from the instrument in
accordance with any applicable safety regulations;
3. Remove the rack(s) from the tray (to open the tray follow the instructions in paragraph
7.7 Load and Unload Sample Vial);
4. Raise the syringe location to the top using the needle motor (4.2.5 “Setup: Manual
operations“);
5. Loosen the left lock;
6. Raise the vial locator;
7. Tighten the left lock;
8. Lift up the sledge manually;
9. Lower the sliding lid;
10. Switch off the autosampler;
11. Unplug the external power supply from the main power and then remove the power cord
from the autosampler;
12. Disconnect the interface cable from the autosampler port and from the analyzer;
13. Disconnect the pneumatic connection;
14. Disconnect, if present, the Ethernet cable from the autosampler and from the Ethernet
connector of the PC or of the Switch;
15. Disconnect, if present, the RS232C cable from the autosampler port and from the serial
port of the PC;
16. Unlock the autosampler from the bracket that fixes it to the analyzer (3.3.1.1 “Unlock the
autosampler from the mounting kit).
17. Raise the autosampler from the bracket and place it in a vertical position.
10.2 Autosampler shipping
The autosampler can only be shipped in the original packaging. If the autosampler has
already been installed you must perform the following:
1. Perform the uninstalling procedure described in paragraph 10.1.