HT2800T_en_c.doc Page 147 of 275
parameters
Air Gap Volume
/
Sample parameters
Volume 1ul
Fill Volume
5ul
Air volume
1ul
Sample Washes
2
Pull Up Strokes
5
Injection parameters
Injection Speed 100ul/sec
Pre Injection Dwell
1sec
Post Injection Dwell
3sec
5.3 Method: SPME mode
5.3.1 Method Menu
This function is used to edit or view the injection method parameters.
Access sequence: RUN>SETTINGS>METHODS
The “RUN: METHODS” screen contains a list of the 10 methods (from 0 to 9) stored in the unit
memory.
The methods can be identified by the number or by a name.
The name can only be created or edited using a PC (using the “HTA Autosampler Manager”
software). Once the names have been created, they will be visible on the touch screen.
Select a method number/name to enter in the “METHOD X” screen (X is a number from 0 to 9
that can be followed by the method name) and to view or edit its parameters. See paragraph
5.3.2 “Method X: Setting method parameters”).
Select Toolsto access the “METHOD TOOLS” screen (see paragra
r
ph 5.3.3 “Method tools “)
TapExitto exit from the “RUN: METHODS” screen.
5.3.2 Method X: Setting method parameters
Each method contains a set of analytical parameters. Use this screen to view or edit these
parameters.
Access sequence: RUN>SETTINGS>METHODS>METHOD X