HT2800T_en_c.doc Page 14 of 275
During Headspace/SPME operation the syringe is cleaned after each injection with inert, clean
gas (e.g. N
2
or He). Therefore a gas supply line must be available close to the autosampler. Do
not use flammable purging gases. Do not use hydrogen as a purging gas. Hydrogen creates a
potential explosion hazard due to the venting of gases.
Only if the Bar code Reader option has been purchase and it is enabled: do not view directly the
led light with optical instruments such as magnifiers, eye loupes or microscopes within a distance
of 100mm. This could cause serious eye injury. The Bar code Reader installed is a Class 1 LED
Product with Class 1 internal radiation. Maximum LED Output: 141,2 uW.
Be careful while handling the syringe holder of the SyringeID system (see 1.7.2 Options” and 11.2
“Liquid mode”). Do not touch the connector or the electronic circuit placed on the back of the
holder. The components may be damaged. If not used, the SyringeID holder must be stored in the
antistatic bag supplied with the packaging.
1.4 Standards Compliance
The applicable rules for this autosampler are the following:
•
2014/35/EU Low-Voltage Equipment Directive.
•
2014/30/EU Electromagnetic Compatibility Directive.
•
2011/65/EU Restriction of the use of certain hazardous substances in electrical and
electronic equipment (RoHS) Directive.
•
2006/42/EC Machinery Directive.
1.5 General Specification
This is an autosampler for gas chromatography.
The HT2800T combines the functions of an autosampler for liquid, static Headspace and SPME
into a single unit.
The autosampler is compatible with most Gas Chromatographs (GC).
It can be installed on GCs with either one or two injectors.In the case of GCs with two injectors,
your autosampler can be configured to inject automatically into one or the other injector.
The autosampler can be programmed by using the touch screen display or by a PC using an
Ethernet or RS232C connection (optional).