HT2800T_en_c.doc Page 220 of 275
Solution:
Tap “ABORT” to stop the operations. Retry the operation.
If the problem is not solved, switch the autosampler OFF, wait 30 seconds and switch it ON
again. Retry the operation.
If the problem persists contact the Technical Assistance Centre, supplying the error message
and error code.
8.15 System integrity test fail
Only for Headspace mode, if the System Integrity tool (see paragraph 1.7.2 “Options“) has been
purchased
Error message:
“The System Integrity test is failed. Please check the integrity of the syringe”
Causes:
The possible causes are:
Septum of the System integrity tool must be replaced.
Syringe (barrel and plunger) must be replaced.
Solution:
• Press RETRY to re-execute the test;
• Press CONTINUE, to go on with the run, even if the test has failed.
• Press ABORT to stop the run and solve the problem.
To solve the problem, the first operation that must be carried out is the replacement of the
septum of the System integrity tool (see paragraph 3.6.3 “System integrity tool installation (only for
Headspace mode)” for the septum replacing instructions. After the septum change, re-execute the
test.
If the System integrity test fails again, press ABORT one more and replace the syringe. Replace
both the syringe barrel and plunger (see paragraph 9.1.5 “Syringe replacement in the
syringe warmer assembly (Headspace mode)”). Re-execute the run.
8.16 Damage to syringe needle
Only for Headspace and Liquid mode. Damaged syringe needles are normally caused by using
incorrect vials, syringes or injection septum; they can also be caused by the incorrect syringe
installation/configuration.
It is recommended to: