8.7 Obstacle found during cover movement
Error message:
Warning: Obstacle found during cover movement!“Try to remove the obstacle
Causes:
The autosampler has found an obstacle during cover movement
Solution:
Remove the obstacle. If necessary, tap “DISENGAGE” to enter the manual operation screen to
move the motors and unlock the system (see paragraph 4.2.5 “Setup: Manual operations).To exit
manual operations, tap “EXIT”.
Tap “RETRY” to re-start the Automatic run procedure from the point of interruption.
8.8 Lack of a Vial (waste or solvent vial for liquid mode,
washing vial for SPME mode)
Error message:
Warning: X missing!
“Insert the vial, then press Retry”
(Where “X” can be the waste vial or a reagent vial)
Causes:
A vial is missing.
Solution:
If you have access to the position, insert the missing vial and tap “RETRY”.
If it is not possible to insert the vial: tap “DISENGAGE” to enter the manual operations screen
and move the motors to allow access to the position and insert the missing vial (see paragraph
4.2.5 Setup: Manual operations); then exit from this screen; the autosampler re-starts
automatically from the point of interruption.
Tapping ABORT” will exit and stop the running operation.