HT2800T_en_c.doc Page 210 of 275
Error message:
“Please add a volume for Fill Volume or set to 0 Pull-up strokes”
Causes:
You have set a “Fill volume” to zero, but you have set “Pull Up Strokes” to a value different to
zero. These settings are not compatible.
Solution:
Tap “CONTINUE” to edit the method (set “Fill volume” to a value different from zero or set the
“Pull Up Strokes” to zero).
8.3.2 Fill volume/Syringe Pre-Fill setting incompatibility
Only for Headspace mode
Error message:
“Please add a volume for Fill Volume or set to NO Syringe Pre-Fill”
Causes:
You have set a “Fill volume” to zero, but you have set the Syringe Pre-Fill to “YES”. These
settings are not compatible.
Solution:
Tap “CONTINUE” to edit the method (set “Fill volume” to a value different from zero or set the
“Syringe Pre-fill” to NO).
8.3.3 Internal Standard setting error
Only for liquid mode
Error message:
“Please add a volume for Internal Standard or disable Internal Standard”
Causes:
You are saving a method, in which in the Internal Standard parameter screen (5.2.2.3 “Internal
standard parameters “) you have set “Enable: YES” but Internal Standard “Volume” to “0”.
Solution:
Tap “CONTINUE”. The Internal Standard parameter screen automatically appears. From this
screen set “Enable: NO” or change the Internal Standard “Volume” to a value different from “0”.
Then tap “EXIT” and save the current method.